Technical data
Digital inputs (1...12)
Factory setting
Voltage range (high)
Voltage range (low)
Max. current per channel (at AC/DC 30 V)
Connection
Maximum cable length
Switching elements
Connection
Number of changeover contacts
Operating principle (changeover contacts)
Function
Electrical endurance under rated operating conditions
Contact data acc. to IEC 60947-5-1
Utilisation category
Rated operational voltage
Rated operational current
Buzzer
Buzzer alarm
Buzzer interval
Buzzer frequency
Buzzer repetition
58
CP305_D00425_02_M_XXEN/07.2023
Can be acknowledged, adoption of characteristics of new
AC/DC 10...30 V
Nominal: 24 V
AC/DC 0...2 V
Plug-in terminals
IN 1... 4 / GND1-4
IN 5...8 / GND 5-8
IN 9...12 / GND 9-12
Plug-in terminal
K1 NC; K1 NO; K1 COM
K2 NC; K2 NO; K2 COM
N/C operation / N/O operation
Programmable
10,000 operating cycles
AC-13
AC-14
AC 24 V
AC 24 V
AC 2 A
AC 2 A
value, can be muted
Configurable
Configurable
Configurable
Off
8 mA
< 500 m
2
DC-12
DC 24 V
AC 2 A