Bender COMTRAXX CP305 Manual page 59

Remote alarm indicator for medical locations and other areas
Hide thumbs Also See for COMTRAXX CP305:
Table of Contents

Advertisement

Device connections
Plug-in terminal (A1/+, A2/-)
Conductor sizes
Stripping length
Rigid/flexible
Flexible with ferrules, with/without plastic sleeve
Multiple conductor flexible with TWIN ferrule with plastic
sleeve
Plug-in terminals
(BMS A, BMS B), (IN1...4, GND1...4, IN9...12, GND9...12) (IN5...8, GND5...8, K1..., K2...)
Conductor sizes
Stripping length
Rigid/flexible
Flexible with ferrule without plastic sleeve
Flexible with ferrule with plastic sleeve
For UL applications
Minimum temperature range of the cable to be
connected to the plug-in terminals
Environment/EMC
EMC
Operating temperature
Operating temperature for UL applications
Operating altitude
Rel. humidity
COMTRAXX® CP305 - Control Panel
IEC 61000-6-2:2016-08 Ed. 3.0
IEC 61000-6-3:2020-07 Ed. 3.0
IEC 61326-1:2020-10 Ed. 3.0
DIN EN 61326-1:2020-10 Ed. 3.0
CP305_D00425_02_M_XXEN/07.2023
AWG 24...12
10 mm
0.2...2.5 mm²
0.25...2.5 mm²
0.5... 1.5 mm²
AWG 18...16
10 mm
0.75...1.5 mm²
0.75... 1.5 mm²
0.75 mm²
Use copper lines only.
75 °C
DIN EN 61326-1:2013-07
DIN EN 50364:2019-05
EN 300 330 V2.1.1
ETSI EN 301 489-3 V2.3.0
-10...+55 °C
-10...+50 °C
≤ 2000 m AMSL
≤ 98 % at 25 °C