Testing Performance
To test the CAL OUTPUT ports (oscilloscope)
Set up the logic analyzer
Set up the Calibration menu.
1
a Press the Waveform key.
b Press the Waveform key again. At the pop up, select Scope Calibration.
c Select the Mode field, then select Service Cal.
d Select the Procedure field, then select DC Cal BNC.
e Select the DC volts field, and set it to O V.
3–66