Testing Performance
To perform the self-tests
For the HP 1660CS-series Logic Analyzers, Select Analy PV, then select Scope PV in
13
the pop-up menu. In the Scope PV menu, select Functional Tests then select All
Tests.
You can run all tests at one time, except for the Data Input Inspection, by running All Tests.
To see more details about each test when troubleshooting failures, you can run each test
individually. This example shows how to run all tests at once.
When the tests finish, the status for each test shows Passed or Failed.
Record the results of the tests on the performance test record at the end of this
14
chapter.
To exit the test system, press the System key, then press the Select key. Select Exit
15
Test, then select Exit Test System.
3–6