Select Sys PV, then select Analy PV in the pop-up menu. In the Analy PV menu,
9
Select All Analyzer Tests.
You can run all tests at one time, except for the Data Input Inspection, by running All
Analyzer Tests. To see more details about each test when troubleshooting failures, you can
run each test individually. This example shows how to run all tests at once.
When the tests finish, the status for each test shows Passed or Failed, and the status for the
All Analyzer Tests changes from Untested to Tested.
Record the results of the tests on the performance test record at the end of this
10
chapter.
For the HP 1660CP-series Logic Analyzers, Select Analy PV, then select Patt Gen in
11
the pop-up menu. In the Patt Gen menu select All Tests.
You can run all tests at one time (except the Output Patterns routine) by selecting All Tests.
To see more details about each test, you can run each test individually. This example shows
how to run all tests at once.
When the tests finish, the status for each test shows Passed or Failed.
Record the results of the tests on the performance test record at the end of this
12
chapter, then continue with step 15.
Testing Performance
To perform the self-tests
3–5