Connect The Logic Analyzer - HP 1660C Series Service Manual

Logic analyzers
Hide thumbs Also See for 1660C Series:
Table of Contents

Advertisement

Set up the Marker menu.
4
a Press the Marker key.
b Move the cursor to the V Markers field and press Select. The voltage markers should
now be On.
c Select Va on C1.
d Select Vb on C1.
e If the T markers are On, turn the T markers Off by moving the cursor to the T markers
field and pressing Select. Select Off.

Connect the logic analyzer

1 Using a BNC-to-banana adapter, connect one end of the cable to the power supply. Connect
the BNC tee, the blocking capacitor, and the shorting endcap to the other end of the cable.
2 Monitor the power supply output with the Digital Multimeter.
To test the voltage measurement accuracy (oscilloscope)
Testing Performance
3–75

Advertisement

Table of Contents
loading

This manual is also suitable for:

1660cs series1660cp series

Table of Contents