Troubleshooting

To verify pattern output (HP 1660CP-Series only)

To verify pattern output (HP 1660CP-Series only)
Equipment Required
Equipment
Oscilloscope
Probe
Output Data Pod
Connect one of the 10460-series data pods to the end of the pattern generator Pod 1
1
cable.
Touch Output Patterns. In the pop-up menu, touch Checkerboard Pattern.
2
Using an oscilloscope, verify the existence of logic-level transitions by touching the
3
oscilloscope probe to each channel of Data Pod 1 and doing an Autoscale. The signal
levels that appear on the oscilloscope display should correspond with the logic
levels represented by the 10460-series pod being used.
5–36
Critical Specification
≥ 500 MHz Bandwidth
500 MHz Bandwidth
no substitute
Recommended Model/Part
HP 54522A
HP 10441A
10460A - series