Set up the Trigger menu.
3
a Press the Trigger key.
b Select the Mode/Arm field, then select Immediate.
Set up the Marker menu.
4
a Press the Marker key.
b Move the cursor to the T Markers field. Press Select, and then press On.
c If the V markers are On, turn the V markers Off by moving the cursor to the V markers
field and pressing Select.

Connect the logic analyzer

Using a BNC-to-banana adapter, connect one end of the cable to the power supply.
1
Connect the BNC tee, the blocking capacitor, and the shorting endcap to the other
end of the cable.
Monitor the power supply output with the Digital Multimeter.
2
Testing Performance
To test the offset accuracy (oscilloscope)
3–79