Testing Performance
To test the input resistance (oscilloscope)

Set up the logic analyzer

Set up the Channel menu.
1
a Press the Config key.
b At the pop up menu, select Scope Channel.
c Select the Input field, then select C1.
d Move the cursor to the Probe field, then use the RPG knob to dial in 1:1.
e Move the cursor to the V/Div field, then use the RPG knob to dial in 20 mV.
f Move the cursor to the Offset field. Set the Offset to 0 V by typing 0, then pressing the
Select key.
g Select the Coupling field, then select 50Ω / DC.
Set up the Trigger menu.
2
a Press the Trigger key.
b Select the Mode/Arm field, then select Immediate.