Set Up The Logic Analyzer; Connect The Logic Analyzer - HP 1660C Series Service Manual

Logic analyzers
Hide thumbs Also See for 1660C Series:
Table of Contents

Advertisement

Set up the oscilloscope.
3
Oscilloscope Setup
Time Base
Display
Time/Div: 1.00 ns/div
mode: avg
delay: 17.7000 ns
# of avg: 16
screens: dual
Channel
Channel 1
on
Display
20.00
Probe Atten
400 mV
Volts/Div
−1.3000 V
Offset

Set up the logic analyzer

Press the Config key. Assign all pod fields to Machine 1. To assign the pod fields,
1
select the pod fields, then select Machine 1 in the pop-up menu.
In the Analyzer 1 box, select the Type field. Select Timing in the pop-up menu.
2

Connect the logic analyzer

Using SMA cables, connect the oscilloscope to the pulse generator channel 1
1
Output, channel 2 Output, and Trig Output.
Using the 6-by-2 test connectors, connect the first combination of logic analyzer
2
clock and data channels listed in the table to the pulse generator.
You will repeat this test for the remaining combinations.
To test the glitch capture (logic analyzer)
Delta V
Delta T
V markers on
T markers on
marker 1 position: Chan 1
start on: Pos Edge 1
marker 2 position: Chan 2
stop on: Pos Edge 1
Channel 2
on
20.00
400 mV
−1.3000 V
Testing Performance
3–19