Set Up The Logic Analyzer - HP 1660C Series Service Manual

Logic analyzers
Hide thumbs Also See for 1660C Series:
Table of Contents

Advertisement

Testing Performance
To test the time measurement accuracy (oscilloscope)

Set up the logic analyzer

Set up the Configuration menu.
1 1
a Press the Config key. At the pop up menu, select Scope Channel.
b Select the Input field, then select C1.
c Move the cursor to the Probe field, then use the RPG knob to dial in 1:1.
d Move the cursor to the V/Div field, then use the PRG knob to dial in 400 mV.
e Move the cursor to the Offset field. Set the offset to 0 by typing 0, then pressing the
Select key.
f Select the Coupling field, then select 50Ω / DC.
g Move the cursor to the s/Div field, then use the RPG knob to dial in 2.00 ns.
Set up the Display menu.
2
a Press the Display key.
b Select the Mode/Arm field, then select Normal.
c Select the Grid field and set it to On.
d In the Waveform selection menu, delete channel 2. If channel 1 is not inserted, insert
channel 1.
3–88

Advertisement

Table of Contents
loading

This manual is also suitable for:

1660cs series1660cp series

Table of Contents