Set up the Trigger menu.
3
a Press the Trigger key.
b Select the Mode/Arm field, then select Edge.
c Select the Source field and set it to C1.
d Move the cursor to the Level field. Set the trigger level to 0 by typing 0 in the
front-panel keyboard, then pressing Select.
Turn off the voltage and time markers.
4
a Press the Marker key.
b Move the cursor to the V Markers field and press Select. The Select key should toggle
the marker to Off.
c Move the cursor to the T Markers field and press Select. At the pop up menu, select
Off.
Testing Performance
To test the trigger sensitivity (oscilloscope)
3–93