Testing Performance
To test the voltage measurement accuracy (oscilloscope)

Set up the logic analyzer

Set up the Channel menu.
1
a Press the Config key. In the pop up menu, select Scope Channel.
b Select the Input field, then select C1.
c Move the cursor to the Probe field, then use the RPG knob to dial in 1:1.
d Select the Coupling field, then select 1MΩ / DC.
e Move the cursor to the s/Div field, then use the RPG knob to dial in 500 ns.
Set the Channel Coupling field to 1MΩ / DC or damage to the equipment will result.
C A U T I O N
Set up the Display menu.
2
a Press the Display key.
b Select the Mode field, then select Average.
c Move the cursor to the Average # field. Type 8 on the front-panel keyboard, then press
Done.
d Select the Grid field and set it to On.
e In the Waveform selection menu, delete channel 2. If channel 1 is not inserted, insert
channel 1.
Set up the Trigger menu.
3
a Press the Trigger key.
b Select the Mode/Arm field, then select Immediate.
3–74