Set Up The Logic Analyzer - HP 1660C Series Service Manual

Logic analyzers
Hide thumbs Also See for 1660C Series:
Table of Contents

Advertisement

Testing Performance
To test the offset accuracy (oscilloscope)

Set up the logic analyzer

Set up the Configuration menu.
1
a Press the Config key. At the pop up menu, select Scope Channel.
b Select the Input field, then select C1.
c Move the cursor to the Probe field, then use the RPG knob to dial in 1:1.
d Move the cursor to the V/Div field, then use the PRG knob to dial in 4.00 V.
e Move the cursor to the Offset field. Set the offset to 0 by typing 0, then pressing the
Select key.
f Select the Coupling field, then select 1MΩ / DC.
Set the Channel Coupling field to 1MΩ / DC or damage to the equipment will result.
C A U T I O N
g Move the cursor to the s/Div field, then use the RPG knob to dial in 500 ns.
Set up the Display menu.
2
a Press the Display key.
b Select the Mode field, then select Average.
c Move the cursor to the Average # field. Type 32 on the front-panel keyboard, then
press Done.
d Select the Grid field and set it to On.
e In the Waveform menu, delete channel 2. If channel 1 is not inserted, insert channel 1.
3–78

Advertisement

Table of Contents
loading

This manual is also suitable for:

1660cs series1660cp series

Table of Contents