Endress+Hauser Proline Cubemass C 100 Operating Instructions Manual page 108

Coriolis flowmeter
Hide thumbs Also See for Proline Cubemass C 100:
Table of Contents

Advertisement

Technical data
108
Mass flow (gases)
±0.50 % o.r.
Density (liquids)
Under reference conditions
[g/cm³]
±0.0005
1)
Valid over the entire temperature and density range
2)
Valid range for special density calibration: 0 to 2 g/cm³, +5 to +80 °C (+41 to +176 °F)
3)
Order code for "Application package", option EE "Special density"
Temperature
±0.5 °C ± 0.005 · T °C (±0.9 °F ± 0.003 · (T – 32) °F)
Zero point stability
DN
[mm]
1
2
4
6
Flow values
Flow values as turndown parameter depending on nominal diameter.
SI units
DN
1:1
[mm]
[kg/h]
1
20
2
100
4
450
6
1 000
US units
DN
1:1
[inch]
[lb/min]
¹⁄₂₄
0.735
¹⁄₁₂
3.675
¹⁄₈
16.54
¹⁄₄
36.75
Standard density calibration
[g/cm³]
±0.02
[in]
¹⁄₂₄
¹⁄₁₂
¹⁄₈
¹⁄₄
1:10
1:20
[kg/h]
[kg/h]
2
1
10
5
45
22.5
100
50
1:10
1:20
[lb/min]
[lb/min]
0.074
0.037
0.368
0.184
1.654
0.827
3.675
1.838
Proline Cubemass C 100 Ethernet/IP
1)
Wide-range
Density specification
[g/cm³]
±0.002
Zero point stability
[kg/h]
[lb/min]
0.0008
0.00003
0.002
0.00007
0.014
0.0005
0.02
0.0007
1:50
1:100
[kg/h]
[kg/h]
0.4
0.2
2
1
9
4.5
20
10
1:50
1:100
[lb/min]
[lb/min]
0.015
0.007
0.074
0.037
0.331
0.165
0.735
0.368
Endress+Hauser
2) 3)
1:500
[kg/h]
0.04
0.2
0.9
2
1:500
[lb/min]
0.001
0.007
0.033
0.074

Advertisement