4.3.2.1 Cable Entry System

The cable entry system is a split system that allows pre-assembled cables to be routed into the protective
housing without disassembling the connectors.
1
Snap-on mounting

Figure 26: Pictures and description of cable entry system

All LAN cables and the 24VDC cable to the Temperature Compensation Sensor Box
must be shielded! Local building codes should be observed when selecting cables!
Longer cable length or less wire gauge causes a certain voltage drop on the power
cable. It has to be ensured, that all devices will be supplied with sufficient voltage
power!
1
Illustrations: © Murrplastik
TIP900
Lay cable into appropriate grommet and provide strain relief
where necessary using cable ties.
Use appropriate tool to punch through cover on base frame.
Set catch hooks into the sides of the cut-out.
Insert rail completely.
The grommet must produce a continuous seal on the back side.
Set the rail onto the catch hooks and press on.
Lock the catch hooks with the rail. Press gently on grommet one
more time.