Brief Operating Instructions Proline Promass 83 Coriolis Mass Flow Measuring System These Brief Operating Instructions are not intended to replace the Operating Instructions provided in the scope of supply. Detailed information is provided in the Operating Instructions and the additional documentation on the CD-ROM supplied.
European standards have been observed. • The manufacturer reserves the right to modify technical data without prior notice. Your Endress+Hauser distributor will supply you with current information and updates to these Operating Instructions. • The information specified on the warning notices, nameplates and connection labels fitted on the measuring device must be observed.
Page 4
"Basic safety requirements" of Appendix I of the Pressure Equipment Directive 97/23/EC. Devices without this identification (without PED) are designed and manufactured according to good engineering practice. • Endress+Hauser will be happy to assist in clarifying any questions on approvals, their application and implementation. Endress+Hauser...
Proline Promass 83 Safety instructions Safety conventions Warning! "Warning" indicates an action or procedure which, if not performed correctly, can result in injury or a safety hazard. Comply strictly with the instructions and proceed with care. " Caution! "Caution" indicates an action or procedure which, if not performed correctly, can result in incorrect operation or destruction of the device.
Installation Proline Promass 83 Installation Transporting to the measuring point • Transport the measuring device to the measuring point in the original packaging. • The covers or caps fitted on the process connections prevent mechanical damage to the sensors during transport and storage. For this reason, do not remove the covers or caps until immediately before installation.
Proline Promass 83 Installation Installation conditions For mechanical reasons, and in order to protect the piping, it is advisable to support heavy sensors. 2.2.1 Dimensions For the dimensions of the measuring device see associated Technical Information on the CD-ROM. 2.2.2 Mounting location The following mounting locations are recommended: •...
Page 8
Installation Proline Promass 83 2.2.3 Orientation • The direction of the arrow on the nameplate of the measuring device must match the flow direction of the fluid. • The following table lists the possible orientations of the measuring devices: Vertical...
Page 9
Proline Promass 83 Installation Special installation instructions for Promass A " Caution! Risk of measuring pipe fracture if sensor installed incorrectly! The sensor may not be installed in a pipe as a freely suspended sensor: • Using the base plate, mount the sensor directly on the floor, the wall or the ceiling.
Installation Proline Promass 83 2.2.6 Inlet and outlet runs No inlet and outlet runs are required. 2.2.7 Vibrations No measures are necessary. Post-installation 2.3.1 Turning the transmitter housing Turning the aluminum field housing Aluminum field housing for non-Ex area £ 180°...
Page 11
Proline Promass 83 Installation Turning the stainless steel field housing £ 180° £ 180° A0007661 max. 360° Release the setscrew. Turn the transmitter housing gently clockwise until the stop (end of the thread). Turn the transmitter Nicht-eigensichere Stromkreise durch IP40-Abdeckung geschützt Non-intrinsically safe counterclockwise (max.
Page 12
Installation Proline Promass 83 2.3.3 Installing the wall-mount housing " Caution! • Make sure that the ambient temperature does not exceed the permitted range. • Always install the wall-mount housing in such a way that the cable entries point downwards.
Wiring Proline Promass 83 Wiring Warning! Risk of electric shock! Components carry dangerous voltages. • Never mount or wire the measuring device while it is connected to the power supply. • Prior to connecting the power supply, connect the protective ground to the ground terminal on the housing.
Proline Promass 83 Wiring Connecting the various housing types Wire the unit using the terminal assignment diagram inside the cover. 3.1.1 Compact version Transmitter connection: Connection diagram inside the connection compartment cover Power supply cable Signal cable or fieldbus cable...
Wiring Proline Promass 83 Degree of protection The devices meet all the requirements for IP 67 (NEMA 4X). After mounting in the field or service work, the following points have to be observed to ensure that IP 67 (NEMA 4X) protection is retained: •...
Proline Promass 83 Hardware settings Hardware settings This section only deals with the hardware settings needed for commissioning. All other settings (e.g. output configuration, write protection, etc.) are described in the associated Operating Instructions on the CD-ROM. Note! No hardware settings are needed for measuring devices with HART or FOUNDATION Fieldbus-type communication.
Page 18
Hardware settings Proline Promass 83 PROFIBUS DP/PA OFF ON OFF ON Device address range: 0 to 126 Factory setting: 126 Miniature switches for device address Example shown: 1+16+32 = device address 49 Miniature switches for the address mode (method of addressing): –...
Proline Promass 83 Hardware settings Device address EtherNet/IP network Has to be set for measuring devices with the communication method: • EtherNet/IP • Miniature switches see description below The IP address can be configured via: • Webserver see section Software settings, "Device address EtherNet/IP network"...
Hardware settings Proline Promass 83 Terminating resistors Note! If the measuring device is used at the end of a bus segment, termination is required. This can be performed in the measuring device by setting the terminating resistors on the I/O board. Generally, however, it is recommended to use an external bus terminator and not perform termination at the measuring device itself.
Proline Promass 83 Commissioning Commissioning Switching on the measuring device On completion of the installation (successful post-installation check), wiring (successful post-connection check) and after making the necessary hardware settings, where applicable, the permitted power supply (see nameplate) can be switched on for the measuring device.
Commissioning Proline Promass 83 Operation 5.2.1 Display elements Display lines/fields Main line for primary measured values Additional line for additional measured +24.502 variables/status variables Information line for bar graph display for example +1863.97 Info icons, e.g. volume flow Current measured values –50...
Proline Promass 83 Commissioning Navigating within the function matrix > 3 s – – – – – – – – A0007665 Enter the function matrix (starting with measured value display) Select the Block (e.g. USER INTERFACE) Confirm selection Select the group (e.g. CONTROL) Confirm selection...
Commissioning Proline Promass 83 Calling the Commissioning Quick Setup All the functions needed for commissioning are called up automatically with the Quick Setup. The functions can be changed and adapted to the process in question. Enter the function matrix (starting with measured value display) Select the group QUICK SETUP...
Proline Promass 83 Commissioning Software settings 5.5.1 Device address PROFIBUS DP/PA, Modbus RS485 Has to be set for measuring devices with the following communication methods: • PROFIBUS DP Device address range 0 to 126, factory setting 126 • Modbus RS485 Device address range 1 to 247, factory setting 247 •...
Page 26
Commissioning Proline Promass 83 5.5.2 Device address EtherNet/IP network Has to be set for measuring devices with the communication method: • EtherNet/IP • Miniature switches see section Hardware settings, The device address can be configured via: "Device address EtherNet/IP network" ä 19 •...
Page 27
Proline Promass 83 Commissioning Displaying the address configurations via the local display > 3 s – … BASIC FUNCTION – … ETHERNET – … CONFIGURATION IP ADDRESS 1 – SUBNETMASK 1 GATEWAY 1 … IP ADDRESS 2 SUPERVISION – SUBNETMASK 2 …...
Commissioning Proline Promass 83 Troubleshooting The error messages that can occur most frequently when a measuring device is commissioned A complete description of all the error messages Operating Instructions on the CD-ROM. are described here. General Note! The output signals (e.g. pulse, frequency) of the measuring device must correspond to the higher-order controller.
Page 29
Proline Promass 83 Commissioning PROFIBUS Device status message PROFIBUS measured value status Extended Cause/remedy (onsite display) diagnostic message in PROFIBUS master • Quality Code (HEX), measured value status: System error mess. (S)/ 0x54; 0x55; 0x56 notice message (!) • Quality status:...