SMC Networks HRR018-A-10 Series Operation Manual page 76

Thermo-chiller
Hide thumbs Also See for HRR018-A-10 Series:
Table of Contents

Advertisement

HRX-OM-W002
Chapter 5 Display and Setting of Various Functions
13.
Press the [▼] key. Display as shown below. It is not used with this product
RUN
14.
Press the [▼] key. Display as shown below. It is not used with this product
RUN
15.
Press the [▼] key. Display as shown below. It is not used with this product
RUN
TEMP READY bandwidth upper limit
16.
Press the [▼] key. Displays TEMP READY bandwidth upper limit setting screen.
RUN
Signal for completion of preparation (TEMP READY)
This function sets the band width for the circulating fluid set temperature (upper/lower
temperature range) in order to notify the customer by communication that the circulating
fluid temperature has reached the band range (upper/lower temperature range).
Alarms can also be assigned.Similar functions include TEMP OUT signal and NOT
TEMP READY signal. Here is an example showing these relationships.
【TEMP READY function setting example】
[1]
Circulating fluid temperature at start of operation: about 20 ° C
[2]
Circulating fluid setting temperature: 15 ° C
[3]
"TEMP READY Bandwidth (Upper Limit)": Set to "2.0 ° C".
"TEMP READY Bandwidth (Lower Limit)": Set to "-2.0 ° C".
[4]
"TEMP READY stabilization time": set to "60 sec".
[5]
"TEMP READY monitoring start time": set to "30 sec".
[6]
"Time when TEMP READY goes off": Set to "10 sec".
[7]
"Switch TEMP READY alarm": Set to "WRN".
[8]
"TEMP READY alarm output": set to "ON".
[9]
"Temperature range upper limit alarm switching": set to "OFF"
[10]
5.5 Advanced setting mode
5-26
Unused
EXT ⇒ HITMP ⇒ LIMIT( Alternately displayed )
Unused
EXT ⇒ LOTMP ⇒ LIMIT( Alternately displayed )
EXT ⇒ TIME( Alternately displayed )
READY bandwidth upper limit ( setting range :0.1 to 9.9 °C )
READY ⇒ HBAND( Alternately displayed )
.
HRR Series