SMC Networks HRR018-A-10 Series Operation Manual page 97

Thermo-chiller
Hide thumbs Also See for HRR018-A-10 Series:
Table of Contents

Advertisement

RS-485 terminal unit
4.
Press the [▼] key. Displays screen for setting RS-485 terminal unit.
Select RS-485 terminal unit.
Slave addresses
5.
Press the [▼] key. Displays screen for setting the slave address.
Sets slave address.
●Setting range: 1 to 32 (Initial value: 1)
Communication speed
6.
Press the [▼] key. Displays screen for setting the communication speed.
Selects the communication speed.
Contact input signal 1 Select function
7.
Press the [▼] key. Displays screen for setting the function of the Contact input signal 1.
HRR Series
RS-485 terminal unit
Set value
Contents
OFF
No terminal
On
With terminal
Slave addresses.
SLV ⇒ ADDR( Alternately displayed )
Communication speed
Set value
Contents
9.6
9600 bps
19.2
19200 bps
Contact input signal 1 function
DIN1 ⇒ SEL( Alternately displayed )
Chapter 5 Display and Setting of Various Functions
Initial value
Initial value
HRX-OM-W002
5.5 Advanced setting mode
5-47

Advertisement

Table of Contents
loading

Table of Contents