Output of NOT TEMP READY
11.
Press the [▼] key. Displays screen for setting the output of the NOT TEMP READY
signal.
RUN
The user can select whether thre is an output signal when alarm "AL13 NOT TEMP READY"
occurs.
Set value
ON
OFF
Example
If the user does not want to generate an alarm, but wants a signal for contact output or serial
communication when alarm "AL13: NOT TEMP READY" occurs, it can be set as follows.
・ Chane NOT TEMP READY alarm :OFF
・ Output of NOT TEMP READY signal:ON
Changing of circulating fluid temperature rise
12.
Press the [▼] key. Displays screen for changing the chiller behaviour when the
circulating fluid temperature rise alarm occurs.
RUN
The user can decide the chiller behavior if alarm "AL14 circulating fluid temperature rise"
occurs.
Set value
OFF
WRN
FLT
HRR Series
Output of NOT TEMP READY signal
TEMP ⇒READY ⇒ OUTP( Alternately displayed )
Contents
Output signal to contact output or serial
communication.
No signal is output.
Operation at occurrence of circulating fluid
temperature rise alarm
TEMP ⇒OUTHI ⇒ ALARM( Alternately displayed )
Contents
Alarm is not generated.
Operation of this product will continue when the
alarm occurs.
Operation of this product will stop when the
alarm occurs.
Chapter 5 Display and Setting of Various Functions
Initial value
No alarm is generated,
but a signal is output.
HRX-OM-W002
Initial value
5.5 Advanced setting mode
5-37