9.6 Replacing Manifold Luer Filters

If the Luer filter fittings become wet and/or contaminated, they can be replaced
easily. First remove the CellASIC
ONIX2 Filter Multiconnector from the front panel,
®
then twist the lock collars counterclockwise to eject the filters. Attach new filters
by placing them loosely on male Luer fittings inside the lock collar. You may need
to adjust the orientation of the filter locking tabs to fit the filter into the lock collar.
Gently twist each lock collar clockwise until the filter is brought up snug inside
the collar. Discard wet/contaminated filters as they cannot be cleaned or reused.
Replacement filters are included in the accessory box that comes with the system,
and can be ordered separately as needed. See Ordering Information.
Figure 44. Luer filter replacement