9.5 Cleaning the Manifold and Tubing

The glass on the top of the CellASIC
gently cleaned with 70% ethanol or a glass cleaner.
NOTE: Manifold components are not autoclavable.
If fluid enters the manifold tubing during the course of an experiment, it should
be cleaned as described below, and the filters should be replaced. A wet filter will
interfere with proper functioning of the system.
NOTE: The manifold cleaning sequence takes approximately 5 minutes to complete,
but may require an overnight air drying step before all tubing is dry and ready for
use in an experiment.
1. Fill the center region of the CellASIC
or hydrogen peroxide (3–6%) up to the max fill indicator line.
CAUTION: DO NOT use ethanol in the Cleaning Plate, as the ethanol will
penetrate the protective filters in the Filter Multiconnector during the cleaning
sequence.
MISE EN GARDE: NE PAS utiliser d'éthanol dans la Plaque de nettoyage car
l'éthanol pénètre dans les filtres de protection du Filtre multi-connecteur au cours
de la séquence de nettoyage.
2. Seal the manifold to the cleaning plate.
Figure 42. Cleaning plate
3. Go to Tools→Cleaning Sequence to open the cleaning sequence interface.
4. Click Clean to begin the cleaning sequence.
20202092 04/19
ONIX2 Manifold XT and Manifold Basic can be
®
ONIX2 Cleaning Plate with distilled water
®
52 of 60