Merck Millipore CellASIC ONIX2 User Manual page 29

Microfluidic system
Table of Contents

Advertisement

7.5 Run Tab,
7.5.3 Dynamic Timeline
G
Figure 17. Run tab during a protocol run
The Dynamic Timeline
activated, two lines will be shown; otherwise there is a single timeline. The movable
scroll line indicates current position in protocol. If dragged left, the past state will be
shown with actual data displayed in the pressure and temperature fields. If dragged
right, the predicted pressure and temperature at that time will be displayed.
Run Details
Status
Run
H
Clock
H
Start Run
Abort
L
Pause
M
Mini-view
20202092 04/19
continued
F
graphically displays the protocol. If dual pressure mode is
F
Displays the current status of the execution of a protocol
G
Changing the selected unit for
H
below it, which display the contents of well groups A1 to D8 on
Plate Layout
the
The elapsed time since the start of the experiment in
dd:hh:mm:ss
The actual 12 hour clock time
Starts the execution of the protocol and begins recording data.
H
If the experiment has already been run, the button will change
to Repeat this experiment, and if clicked will create a new
experiment based on the currently open experiment. Repeat
this experiment is same as if you clicked New Experiment
Based on Previous and selected the file of the currently open
experiment. The data of the previous experiment is unmodified
and a new experiment file is created, populated with everything
from the previous experiment except for the run data.
Aborts the execution of the protocol; data collected so far is
saved in the experiment file
Temporarily pauses execution of the protocol. While paused, the
button changes to
when the user desires.
temperature and pressure settings. When the user resumes the
experiment, previous well groups and pressure configuration
are restored. The
whenever the protocol hits a programmed undetermined pause.
However, determined pauses (e.g., a 5 minute programmed
pause) cannot be manually resumed before the end of their
scheduled duration.
Opens the floating mini-view dynamic timeline
N
H
I
J
Status
tab.
Resume
and experiment can be resumed
Pause
turns off well groups but keeps
Pause
button also changes to
25 of 60
K
L
updates the text fields
Resume
CellASIC
ONIX2 Microfluidic System
®
N
M

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Millipore CellASIC ONIX2 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

This manual is also suitable for:

Cax2-s0000

Table of Contents