5. 2. 5 Manual wipe-down disinfection
NOTE: To prevent liquid entering the device, it should not be sprayed with the reprocessing agent.
1
NOTE: In order to preserve the conductivity of the contacts, you should never allow reprocessing
1
agents to come into contact with the connections on the front and back of the device. When the
cleaning agent dries, residues can form which impair conductivity.
1. Use a disposable cloth moistened with disinfectant or a soaked ready-to-use disinfectant cloth.
2. Wipe-clean the exterior surfaces of the medical device.
3. Allow the disinfectant to take effect, observing the instructions of the disinfectant manufacturer.
4. At the end of the necessary exposure time, wipe the surfaces with a dry low lint cloth.
5. After using the cloth, dispose of this in accordance with the disposal regulations.
5. 3 Maintenance and repair
5. 3. 1 Maintenance plan
Interval
As required
Annually
Every 4 years
5. 3. 2 Replacing the BIOS battery
The BIOS battery must be replaced by KARL STORZ service technicians or appropriately trained
specialist personnel.
5. 3. 3 Changing the device fuse
The device fuse must be changed by KARL STORZ service technicians or appropriately trained
specialist personnel.
5. 3. 4 Servicing
Defective device components must be serviced and repaired exclusively by persons authorized by
KARL STORZ; all repair work must employ original KARL STORZ parts only.
All service work may only be performed by KARL STORZ or by an authorized agent.
5-2
Action
Change fuses
Check and inspect device
Replace BIOS battery
Maintenance
Responsibility
KARL STORZ service technician
KARL STORZ service technician
KARL STORZ service technician
96296083EN / BA