Karl Storz AIDA WD 300 System Description page 23

Table of Contents

Advertisement

3
Installation
3. 1 Unpacking
Check the delivery for missing items and evidence of shipping damage. Please file any complaints with
the manufacturer or supplier immediately.
Retain the original packaging for later use. This may be useful if the device has to be transported.
3. 2 Installation and commissioning
WARNING: Installation and commissioning of the KARL STORZ AIDA
3
by authorized, electrotechnically qualified specialists from KARL STORZ GmbH & Co. KG or by an
authorized contractual partner.
WARNING: Combinations of medical devices are only assured to be safe if they are identified as such
3
in the respective instruction manuals or if the intended use and interface specifications of the products
used in combination permit this.
WARNING: The KARL STORZ AIDA
3
ventilation.
WARNING: Connect the potential equalization cable to the potential equalization connector. The
3
device's ground line should be installed by a qualified electrician.
WARNING: The electrical installations in the operating room in which the device is installed and
3
operated must comply with the applicable IEC standards.
KARL STORZ AIDA
®
devices must be installed free-standing so as to ensure sufficient
®
Installation
must only be performed
®
3-1