Karl Storz AIDA WD 300 System Description page 103

Table of Contents

Advertisement

'
'
DICOM
menu
Set up the DICOM interface in this menu.
NOTE: Consult with your facility's IT department as regards setting up the DICOM interface.
1
Setting up DICOM parameters
1. Under Client AET, enter the AE title which was assigned to the KARL STORZ AIDA
your IT department.
2. Under Modality, select a Modality Type.
ES: Endoscopy
US: Ultrasound
OT: Other
3. Under Char. set, select the character set in collaboration with your IT department.
4. The port describes the channel on which the DICOM service of the AIDA application is running.
This is required for the Storage Commitment and in order to check the connection between the
DICOM server and the AIDA application.
5. Pressing the Encryption button will take you to a selection screen where you can choose
between two encryption protocols for data transfer; ISCL (Internet Security Consulting Language)
or TLS (Transport Layer Security)
If no encryption protocol is selected, data will be transferred without any encryption.
6. If a check mark is placed next to 'Local Cert. file', the local computer's certificates for TLS have
been imported on this computer. These are imported by pressing the Browser button, and are
usually imported from a USB stick. The certificates are managed by the hospital's IT department.
7. If a check mark is placed next to 'Server Cert. file', certificates have been loaded for the DICOM
server used by the application, also via the Browser function. These server certificates are
managed by the hospital's IT department.
8. Leave the LogLevel as the factory setting.
NOTE: The LogLevel can be adjusted for service purposes, but should normally remain in factory
1
settings.
KARL STORZ AIDA
®
Description of the program functions
system by
®
2-11