Table 203
Test specification for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment
(according to IEC 60601-1-2)
The KARL STORZ AIDA
The user of the KARL STORZ AIDA
Test
Band
frequency
MHz
MHz
385
380 bis 390
450
430 bis 470
710
704 bis 787
745
LTE
780
810
870
800 bis 960
930
1720
1845
1700 bis
1970
1990
2450
2400 bis
2570
5240
5100 bis
5500
5800
5785
* NOTE: If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and
the KARL STORZ AIDA
3.
WD 300
KARL STORZ AIDA
®
®
WD 300 is intended for use in the electromagnetic environment specified below.
®
WD 300 should make sure that they are used in such an environment.
Service
TETRA 400
GMRS 460,
FRS 460
Band 13, 17
GSM 800/900
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
GSM 1800,
CDMA 1900,
GSM 1900,
DECT,
LTE Band 1, 3,
4, 25, UMTS
Bluetooth,
WLAN 802.11
b/g/n,
RFID 2450,
LTE Band 7
WLAN 802.11
a/n
®
WD 300 may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-
Modulation
Maximum
Power
Puls
1,8
modulation
18 Hz
FM
± 5kHz Hub
1 kHz Sinus
Puls modulatio
0,2
217 Hz
Puls modulatio
18 Hz
Puls modulatio
217 Hz
Puls modulatio
217 Hz
Puls modulatio
0,2
217 Hz
Appendix
Distance
TEST LEVEL
m
W
0,3
2
0,3
0,3
2
0,3
2
0,3
2
0,3
0,3
IMMUNITY
V/m
27
28
9
28
28
28
9
7-5