Karl Storz AIDA WD 300 System Description page 104

Table of Contents

Advertisement

Setting up the interface to the DICOM Worklist provider or MPPS
1. If you want to allow access to a DICOM Worklist or MPPS, place a check mark next to Enable
Worklist or Enable MPPS.
NOTE: If you remove the check mark, the Import DICOM Worklist or MPPS button will no longer be
1
displayed on the user interface.
NOTE: Do not make the settings described below when you have removed the check mark.
1
NOTE: The fields Server AET, Server IP and Port are outlined in yellow until an entry has been
1
made.
2. Under Server AET, enter the AE title of the DICOM Worklist server.
3. Under Server IP, enter the IP address of the DICOM Worklist server.
4. Under Port, enter the port for the DICOM Worklist server.
NOTE: With MPPS, the option to select different date display options is not available.
1
5. Under Date filter, limit the date range which is to be displayed when searching for patient data in
the DICOM Worklist.
Today: Only display the patients on the Worklist for the current day.
One Day: plus/minus 1 day
Two Days: plus/minus 2 days
Three Days: plus/minus 3 days
Select None if you do not wish to limit the date range.
6. Set the Timeout.
Timeout:
NOTE: The Timeout can be changed if necessary, but should otherwise remain in factory settings.
1
7. Tap on Echo to test the connection to the DICOM server or MPPS.
NOTE: If the connection test fails, there is probably an error in the network connection or in the
1
entered server access data.
NOTE: Consult with your facility's IT department as regards connection problems.
1
2-12
30 seconds
Description of the program functions
96296082EN / BA

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Aida wd 350

Table of Contents