Karl Storz AIDA WD 300 System Description page 107

Table of Contents

Advertisement

'
'
HL7
menu
Set up the HL7 interface in this menu.
Two separate HL7 server settings can be configured.
• Query patient data from HL7 configuration, with a choice between DEM and APA
• HL7 export settings configuration for the storage of ORU and MDM messages.
Both server settings have an echo function, meaning that the server availability can be checked
immediately after configuration has been carried out.
Setting up HL7 parameters
1. Under HIS name, enter the name of the hospital information system.
2. Under Modality name, select the name for the AIDA system allocated to you by the IT
department.
3. Under PID field, select the position you would like the patient ID to appear within the HL7
messages.
4. Under Version, select the version of the HL7 standard currently in use.
Setting up an interface to the HL7 server
1. If you wish to allow a request for patient data, place a check mark next to Allow HL7 request.
2. Under Server IP, enter the IP address of the HL7 server.
3. Under Port, enter the port for the HL7 server.
4. Set the Max. time for answer.
Time for Answer:
5. Under Value QRD-9, select the patient search type.
DEM:
APN:
KARL STORZ AIDA
®
60 seconds
Search according to patient ID
Search according to case number
Description of the program functions
2-15