Internal failure

Informational tones

Alarm history

7-4
Red indicates less than 5 minutes.
'Internal problem prevents normal operation.' shows on the display
during a software or hardware failure that requires service. If this
message occurs, contact an authorized service representative.
The system provides informational tones that are 43 to 78 dB(A)
depending on the alarm volume setting.
Notification tone is a medium pitched tone that sounds for
approximately 480 milliseconds. This tone signifies that an
action must be taken.
Reject tone is a low pitched tone that sounds for approximately
200 milliseconds. This tone signifies that the previous action
was rejected.
Use Alarm History to view a list of the 12 most recent high and
medium alarms that occurred since system power up. The most
recent alarm shows on the top of the list. If more than 12 alarms
have occurred, the oldest alarms are deleted as new ones are
logged.
The alarm history clears when the system is shut down or when a
new case is started. All alarm information and operation steps are
saved in service log. All logs can be accessed using in Super User
mode or service mode.