Power - Respironics BiPAP Focus Operator's Manual

Noninvasive ventilator system
Hide thumbs Also See for BiPAP Focus:
Table of Contents

Advertisement

Power

Input range
Battery backup
Nurse call station
relay
Fuse type
Power cord
®
BiPAP
Focus™ Noninvasive Ventilator System
1027404 Rev K
~
100-240 V
, 50/60 Hz, 3.0-1.5 A
3.8 amp-hour (Ah) nickel metal hydride (NiMH) battery
provides a minimum of 45 minutes of operation at
default settings. System automatically activates the
battery charge cycle when connected to AC power and
the ON/OFF switch is ON ( | ) (during normal operation
or Standby mode).
Recharge time is typically under 5 hours, but may
extend further depending on machine settings if
operating on battery prior to recharge, or if operating in
elevated ambient temperature (above 28°C). Charge is
complete when the green CHARGING LED
extinguishes.
The nurse call station port is intended to connect only to
a SELV (Safety Extra-Low Voltage) and ungrounded
system with basic insulation to ground, in accordance
with IEC 60601-1. To prevent damage to the nurse call
circuit, the signal input should not exceed the maximum
rating of 24 VAC (Volts of Alternating Current) or 36
VDC (Volts of Direct Current) at 500 mA (milliamps),
with a minimum current of 1 mA.
No replaceable fuses.
Use the BiPAP Focus System only with the following
power cords:
Europe: Plug type CEE 7/7
(P/N 1029985)
UK/Ireland: Plug type BS 1363
(P/N 1029982)
USA/Canada: Plug type NEMA 5-15
(P/N 1029989)
Contact your local distributor for other countries.
Specifications
73