The BiPAP Focus System is suitable for use in the specified
electromagnetic environment. The customer and/or the user of the BiPAP
Focus System should ensure that it is used in an electromagnetic
environment as described below:
Immunity Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
NOTE: UT is the AC mains voltage prior to application of the test level.
BiPAP® Focus™ Noninvasive Ventilator System
1027404 Rev K
IEC 60601
Compliance
Test Level
Level
± 6 kV
± 6 kV
contact
contact
± 8 kV air
± 8 kV air
± 2 kV for
± 2 kV for
power
power
supply lines
supply lines
± 1 kV for
± 1 kV for
input/output
input/output
lines
lines
± 1 kV
± 1 kV
differential
differential
mode
mode
± 2 kV
± 2 kV
common
common
mode
mode
Electromagnetic
Environment Guidance
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Mains power quality should
be that of a typical
commercial and/or hospital
environment.
Mains power quality should
be that of a typical
commercial and/or hospital
environment.