Alarm Message
High
Temperature
Key Stuck
Lo P Reg
(no auto-reset)
Loss of AC
(auto-resets)
®
BiPAP
Focus™ Noninvasive Ventilator System
1027404 Rev K
Table 4: BiPAP Focus Alarms
Description
High temperature detected
inside the ventilator.
One or more of the user
interface keys has been
detected as permanently
pressed or stuck closed.
Low pressure regulation
alarm. Circuit pressure is 5
cmH
O below set CPAP or
2
IPAP for 60 sec.
Ventilator has switched to
battery operation. This is a
notification alarm.
Auto-resets when AC power
is connected.
After the Alarm Reset is
pressed the unit beeps
once a minute when on
battery.
Corrective Action
Check room temperature is
within specified operating
conditions. Allow the
ventilator to acclimate to the
ambient conditions.
Check the keypad.
Run pre-use check.
Contact customer service.
Check patient and settings.
Check for leaks in breathing
circuit or mask fitting.
Press ALARM RESET to
clear.
If not in use for transport,
check connection to AC
power and verify that AC
POWER and green
BATTERY STATUS
indicators are lit.
Alarms
51