Measured Data - Respironics BiPAP Focus Operator's Manual

Noninvasive ventilator system
Hide thumbs Also See for BiPAP Focus:
Table of Contents

Advertisement

Settings: S/T Mode
I-Time
Inspiratory time
Ramp Time
Ramp Start
Initial
inspiratory
pressure

Measured Data

Patient circuit
pressure bar
graph
(continuous
display)
Rate
Measured
respiratory rate
®
BiPAP
Focus™ Noninvasive Ventilator System
1027404 Rev K
Range: 0.5-3 sec
Resolution: 0.1 sec
Accuracy: ± (0.1 + 10% of setting) sec
I-Time and Rate settings cannot allow I-Time to exceed
expiratory time.
Range: 0-45 min
Resolution: 5 min
Accuracy: ±10% of setting
Range: 4 cmH
O to EPAP (4 hPa to EPAP)
2
Resolution: 1 cmH
Range: 0-35 cmH
Resolution: 1 cmH
Accuracy: ±10% of scale
In the event of total loss of power, the inspiratory and
expiratory pressure measured at the patient exhalation port
at 60 L/min is less than 1 cmH
ventilator design.
Range: 0-60/min
Resolution: 1/min
Accuracy: ± (1 + 10% reading)
O (1 hPa)
2
O (0-35 hPa)
2
O (1hPa)
2
O by virtue of mask and
2
Specifications
69