drop due to the BiPAP Vision single patient use circuit and the
bacteria filter (see "Accessories" on page 87 for ordering
information).
WARNING: Check that the orientation of the oxygen safety valve is
correct.
WARNING: Connect oxygen only on order of a physician.
WARNING: When connecting oxygen to the circuit, use an in-line
safety valve and do not allow oxygen to be entrained through the air
inlet.
WARNING: Observe the patient and monitor continuously using
pulse oximetry and arterial blood gas measurements as needed
during oxygen administration.
WARNING: Inspired oxygen concentration can vary according to
settings, patient breathing patterns, and leak rate.
WARNING: To avoid oxygen accumulation in the ventilator, put it
into Normal operating mode before turning the oxygen ON, and turn
oxygen OFF before putting it into Standby or turning the unit OFF.
WARNING: Do not use a Heat and Moisture Exchanger (HME) with
the BiPAP Focus due to high resistance to flow.
WARNING: The BiPAP Focus System can deliver pressures up to 30
cmH
O. In the unlikely event of certain fault conditions a maximum
2
static pressure of 42 cmH
WARNING: It is important to inspect and replace the air inlet filters at
the recommended intervals to avoid introducing foreign matter into
the system
WARNING: The BiPAP Focus System is shipped with the battery
disconnected. The battery must be connected before the BiPAP
Focus System will operate.
WARNING: To avoid the risk of fire, use only Respironics-approved
batteries. All battery connections are keyed to ensure proper
connection.
®
BiPAP
Focus™ Noninvasive Ventilator System
1027404 Rev K
O is possible.