Power; Triggers And Cycles; Environmental - Respironics BiPAP Focus Service Manual

Ventilator
Hide thumbs Also See for BiPAP Focus:
Table of Contents

Advertisement

REF 1029568 Rev A

Power

Input range
Battery backup
Nurse call station relay
Fuse type
Power cord

Triggers and Cycles

Volume-based trigger
Flow reversal cycle
Shape signal
Timed trigger
IPAP maximum of 3.0 seconds
Peak flow cycle
Supplemental Oxygen
Flow and pressure into
oxygen valve

Environmental

Temperature
Atmospheric pressure
BiPAP® Focus™ Ventilator Service Manual © Respironics, Inc.
90-264 V~, 47-63 Hz, 90 VA
3.8 amp-hour (Ah) nickel metal hydride (NiMH) battery
provides 45 minutes of operation at default settings.
System automatically activates the battery charge cycle
when connected to AC power and the ON/OFF switch is
ON ( | ) (during normal operation or Standby mode).
Recharge time is typically under 5 hours, but may extend
further depending on machine settings if operating on
battery prior to recharge, or if operating in elevated
ambient temperature (above 28°C). Charge is complete
when the green CHARGING LED extinguishes.
Rated current: 0.280 A
Rated voltage: 250V
No replaceable fuses.
Refer to the"Complete Parts List" on page 9-1
Maximum flow: 15 L/min at ambient pressure
Operating: +5 to +35°C at 10 to 95% relative
humidity
Storage: -20 to 60°C at 95% relative humidity
Operating: 83 to 102 kPa (830 to 1020 mBar)
Chapter 3
Theory of Operation
3-11