Respironics BiPAP Focus Operator's Manual page 64

Noninvasive ventilator system
Hide thumbs Also See for BiPAP Focus:
Table of Contents

Advertisement

Care and Maintenance
Table 5: BiPAP Focus System Maintenance
Part
Air inlet
filter(s)
Internal valve
assembly
System
exterior
58
Interval
Reusable: clean
monthly or as
needed.
Reusable: every
year or as needed.
Disposable: replace
as needed.
NOTE: The air path
cannot be cleaned.
Every 10,000 hours
of operation.
Refer to the BiPAP
Focus Service
Manual to check
hours of operation.
As needed.
BiPAP
Procedure
Follow the "Cleaning the Reusable Air
Inlet Filter" on page 8.
Replace.
Do not attempt to clean or reuse
disposable filters.
Qualified service technician only.
Clean the exterior of the ventilator with
a soft damp cloth moistened with any of
the following solutions:
• Mild detergent or soapy water
• 10% bleach solution (90% water)
• Isopropyl alcohol (91%)
• Quaternary ammonium
germicides (sprays or disposable
cloths)
Do not spray or immerse in liquid. Do
not allow liquid to penetrate the system.
®
Focus™ Noninvasive Ventilator System
1027404 Rev K