Sample FiO
values for different tidal volumes with RR=20 I:E = 1:2 10L/
2
min patient leak, and using a Vision Circuit with IPAP/EPAP settings of
10/5,20/5,30/5.
Tidal Volume
(mL)
600
1000
WARNING: Connect oxygen only on order of a physician.
WARNING: When connecting oxygen to the circuit, use an in-line
safety valve and do not allow oxygen to be entrained through the air
inlet.
WARNING: Observe the patient and monitor continuously using
pulse oximetry and arterial blood gas measurements as needed
during oxygen administration.
WARNING: Inspired oxygen concentration can vary according to
settings, patient breathing patterns, and leak rate.
WARNING: To avoid oxygen accumulation in the ventilator, put it
into Normal operating mode before turning the oxygen ON, and turn
oxygen OFF before putting it into Standby or turning the unit OFF.
WARNING: To avoid the risk of fire, use this device in well-ventilated
areas away from flammable anesthetics. Do not use in a hyperbaric
chamber or other similarly oxygen-enriched environments.
®
BiPAP
Focus™ Noninvasive Ventilator System
1027404 Rev K
Table 1: Oxygen Concentration
FiO
@ 5 L/min
2
Supplemental Flow
37 +/- 3%
30 +/- 3%
FiO
@ 10 L/min
2
Supplemental Flow
50 +/- 3%
40 +/- 3%
Setup