Recommended separation distances between portable and mobile RF
communications equipment and the BiPAP Focus IEC 60601-1-2:
Frequency of
transmitter
Equation
Rated maximum
output power of
transmitter (W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the separation
distance can be estimated using the equation in the corresponding column, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.
BiPAP® Focus™ Noninvasive Ventilator System
1027404 Rev K
150 kHz to
80 MHz
P
d = 1.2
Separation distance
(m)
0.12
0.38
1.2
3.8
12
150 kHz to
800 MHz
P
d = 1.2
Separation distance
(m)
0.12
0.38
1.2
3.8
12
Compliance
800 MHz to 2.5
GHz
P
d = 2.3
Separation distance
(m)
0.23
0.73
2.3
7.3
23
85