Maximum Pressure Drop for Disconnect

The Disconnect alarm relies on a fixed relationship between the patient
pressure settings and the open circuit flow of the Vision
use circuit. The alarm works correctly unless the circuit is excessively
restrictive. Use the pre-operational check to confirm that the Disconnect
alarm operates properly with the circuit.

Triggers and Cycles

Volume-based trigger
Flow reversal cycle
Shape signal
Timed trigger
IPAP maximum of 3.0 seconds
Peak flow cycle

Supplemental Oxygen

Flow and pressure into
oxygen valve

Environmental

Temperature
Atmospheric pressure
Storage
®
BiPAP
Focus™ Noninvasive Ventilator System
1027404 Rev K
Maximum flow: 15 L/min at ambient pressure
Operating: +5 to +35°C at 10 to 95% relative
humidity
Storage: -20 to 60°C at 95% relative humidity
Operating: 83 to 102 kPa (830 to 1020 mBar)
When the BiPAP Focus is not in use for periods
of 2 weeks or longer, it is recommended that the
unit be stored in a clean, sanitized plastic bag.
Specifications
®
single patient
75