Alarms
Alarm Message
Battery Depleted
(auto-resets)
Disconnect
(auto-resets)
Hi P Reg
(no auto-reset)
High Batt Temp
(auto-resets)
50
Table 4: BiPAP Focus Alarms
Description
Up to 2 minutes of battery
power remain.
Auto-resets when AC power
is connected.
Cannot be manually reset
or silenced.
Ventilator has detected a
patient disconnect condition
for at least 15 seconds.
Auto-resets when
disconnect condition no
longer exists.
High pressure regulation
alarm. Circuit pressure 5
cmH
O above set CPAP or
2
IPAP for 500 msec.
Battery is overheating.
Battery may be almost
depleted; or there may be a
patient or breathing circuit
disconnect, causing the
blower to run at high speed.
Auto-resets when AC power
is connected.
BiPAP
Corrective Action
Check connection to AC
power and verify that AC
POWER and green
BATTERY STATUS
indicators are lit.
If battery was depleted
earlier than expected, refer
to "Testing the Battery" on
page 61.
Replace battery if
necessary.
Check patient.
Check breathing circuit
connections.
Check patient and settings.
Verify that breathing circuit
is not obstructed.
Check patient.
Connect ventilator to AC
power.
Check for causes of
overheating, including a
patient or breathing circuit
disconnect.
®
Focus™ Noninvasive Ventilator System
1027404 Rev K