Settings: S/T Mode - Respironics BiPAP Focus Operator's Manual

Noninvasive ventilator system
Hide thumbs Also See for BiPAP Focus:
Table of Contents

Advertisement

Specifications

Settings: S/T Mode

IPAP
Inspiratory
positive airway
pressure, the
inspiration
pressure setting
EPAP
Expiratory
positive airway
pressure, the
exhalation
pressure setting
Rise-Time
How quickly the
ventilator
increases
inspiratory
pressure from
EPAP pressure
to 67% of
pressure
support level
Rate
Respiratory
rate, used to
determine if a
timed breath is
delivered
68
Range: 4-30 cmH
Resolution: 1 cmH
Dynamic accuracy: ±5 cmH
IPAP cannot be set below EPAP.
Range: 4-25 cmH
Resolution: 1 cmH
Dynamic accuracy: ±5 cmH
Range: 1-6 (where 1 = 0.1 sec and 6 = 0.6 sec)
Resolution: 1
Accuracy: ± (0.15 + 10% of setting) sec
Range: 1-30/min
Resolution: 1/min
Accuracy: ±1/min or ±10% of setting, whichever is greater
over a 4-minute period.
I-Time and Rate settings cannot allow I-Time to exceed
expiratory time.
BiPAP
O (4-30 hPa)
2
O (1 hPa)
2
O (5 hPa)
2
O (4-25 hPa)
2
O (1 hPa)
2
O (5 hPa)
2
®