8. If the blower is on, turn it off as follows:
a. Scroll using the DOWN arrow key until System Test is
highlighted, and then press ENTER. The System Test
screen shown in Figure 4 will appear.
b. Scroll using the DOWN arrow key until the Blower status is
highlighted, and then press ENTER to toggle it Off.
c. Press CHANGE SCREEN to return to the Menu screen.
9. Disconnect the ventilator from AC power.
10. Record the VBATT reading with the blower off, as follows:
a. From the Menu screen, scroll using the arrow keys until
Sensor Readings is highlighted, and then press ENTER.
The Sensor Readings screen will appear (Figure 5).
®
BiPAP
Focus™ Noninvasive Ventilator System
1027404 Rev K
Figure 33: System Test Screen
Figure 34: Sensor Readings Screen
Care and Maintenance
63