Alarm Settings - Respironics BiPAP Focus Operator's Manual

Noninvasive ventilator system
Hide thumbs Also See for BiPAP Focus:
Table of Contents

Advertisement

Specifications
Measured Data
Est. Vt
Estimated
delivered tidal
volume
Est. MV
Estimated
exhaled minute
volume
Leak
Estimated
Patient leak
#Apnea
Brief apnea
periods
Trigger
Breath trigger
NOTE: Est. Vt and Est. MV estimates are at ambient temperature and
pressure, dry (ATPD). Pressure signals are filtered using a 50-Hz low pass
Butterworth filter.

Alarm Settings

Apnea
The length of time without a spontaneous
breath that triggers the Apnea alarm
70
Range: 0-4000 mL
Resolution: 1 mL
Accuracy (S/T Mode): ± (50 mL + 10% reading) (when leak
<60 L/min, using the Vision circuit)
Accuracy (CPAP Mode): ± (100 mL + 10% reading) (when
leak <60 L/min, using the Vision circuit)
Vt display flashes when peak inspiratory flow for successive
breaths varies by more than 15 L/min.
Range: 0-99 L/min
Resolution: 0.1 L/min
Accuracy: ± 1 L or ± 10% actual, whichever is greater (when
leak is <60 L/min using the Vision circuit).
Range: 0-150 L/min
Resolution: 1 L/min
Accuracy: ± (15 L/min +10%)
Range: 0-99/hour
Resolution: 1/hour
Accuracy: ± 1/hour (after 1 hour)
Range: Patient or Timed (ventilator)
BiPAP
Range: 20, 40, 60 sec or
OFF
Default setting: 20 sec
®
Focus™ Noninvasive Ventilator System
1027404 Rev K