Respironics BiPAP Focus Operator's Manual page 58

Noninvasive ventilator system
Hide thumbs Also See for BiPAP Focus:
Table of Contents

Advertisement

Alarms
Alarm Message
Low Battery
(auto-resets)
Low Temperature
System Error:
XXX
(no auto-reset)
Any alarm not
shown above
52
Table 4: BiPAP Focus Alarms
Description
Approximately 10 minutes
of battery power remain.
Auto-resets when AC power
is connected.
Low temperature detected
inside the ventilator.
The system has halted and
is not providing breathing
support to the patient.
BiPAP
Corrective Action
Check connection to AC
power and verify that AC
POWER and green
BATTERY STATUS
indicators are lit.
If battery was depleted
earlier than expected, refer
to "Testing the Battery" on
page 61.
Replace battery if
necessary.
Check room temperature is
within specified operating
conditions. Allow the
ventilator to acclimate to the
ambient conditions.
Check patient and provide
alternate breathing support.
Cycle power: if problem
persists, remove ventilator
from use and contact a
qualified service technician.
Contact customer service.
®
Focus™ Noninvasive Ventilator System
1027404 Rev K