Removing a Cassette
WARNING: Close the fluid path tubing with the clamp before
removing the cassette from the pump to prevent unregulated gravity
infusion, which could result in death or serious injury to the patient.
1. Stop the pump.
2. Close the tubing clamp.
3. Insert a coin into the lock and turn it clockwise. The lock
will pop out when you unlock the cassette.
4. A continuous alarm will sound and the pump will display
"No Disposable, Clamp Tubing." The alarm may be si-
lenced by pressing
5. Remove the cassette hooks from the pump hinge pins.
Section 2: Pump Setup and Programming
⁄
„
or
.
31