Section 2: Pump Setup and Programming
WARNING: When the Air Detector is turned off, the pump
will not detect air in the fluid path. Periodically inspect the
fluid path and remove any air to prevent air embolism. Air
embolism could result in death or serious injury to the
patient.
• If you need to change the Air Detector setting, see Section 4,
Biomed Functions.
„
• Press
11. Verify the Upstream Sensor status.
• Make sure the desired setting is displayed. This screen will
show whether the Upstream Occlusion Sensor is turned on
or off.
WARNING: When the Upstream Occlusion Sensor is turned
off, the pump will not detect occlusions upstream (between
pump and fluid container). Periodically inspect the fluid
container for decreasing volume, inspect the fluid path for
kinks, a closed clamp, or other upstream obstructions.
Upstream occlusions could result in under- or non-delivery
of medications. If undetected, these occlusions could result
in death or serious injury to the patient.
• If you need to change the Upstream Occlusion Sensor
setting, see Section 4, Biomed Functions.
„
• Press
12. Review the program.
„
Press
repeatedly to review the programming screens. If
you need to reprogram a setting, press
ate screen appears and change the setting as described in this
section.
30
.
.
„
until the appropri-