Section 1: General Description
occlusions could result in death or serious injury to the patient.
Downstream Occlusion Sensor: The pump contains a downstream
occlusion sensor. When a downstream occlusion (between the pump
and patient access site) is detected, an alarm will sound, delivery
will stop, and the display will show "High Pressure."
Reservoir Volume Alarm: The Reservoir Volume alarm indicates
when the fluid in the fluid container is low or depleted. Each time
you change the fluid container, you may reset the Reservoir Volume
to the originally programmed volume. Then, as medication is
delivered, the Reservoir Volume automatically decreases. When the
pump calculates that 5 ml remain in the fluid container, beeps
sound and "ResVol Low" appears on the main screen. This alarm
recurs at every subsequent decrease of 1 ml until the Reservoir
Volume reaches 0 ml, at which point the pump stops and the
Reservoir Volume empty alarm sounds.

The Main Screen

The main screen is the starting point for programming or viewing
the pump's settings.
If no keys are pressed for a period of time (2 minutes), the display
reverts to the main screen. When the two AA batteries are low,
"LowBat" appears on the main screen.
When running:
Status of pump
Reservoir Volume
When stopped:
Status of pump
8
RUN
ResVol
50.0 ml
STOPPED
Battery Status
Status of Reservoir
LowBat