Upstream Sensor On/Off

The Upstream Occlusion Sensor screen can be set to On or Off. If this
screen is set to On, and an upstream occlusion (between pump and
fluid container) is detected, an alarm will sound, delivery will stop,
and the display will show "Upstream Occlusion."
WARNING: When the Upstream Occlusion Sensor is turned off,
the pump will not detect occlusions upstream (between pump and
fluid container). Periodically inspect the fluid container for decreas-
ing volume, inspect the fluid path for kinks, a closed clamp, or
other upstream obstructions. Upstream occlusions could result in
under- or non-delivery of medications. If undetected, these occlu-
sions could result in death or serious injury to the patient.
1. With the pump stopped and in LL0, access Biomed Func-
tions. (Refer to the beginning of the Biomed Functions
section for instructions on how to access Biomed Functions.)
2. Press
´
3. Press
¤
4. Press
until "Upstream Sensor" appears.
Î
or
to select Off or On.
to enter the change.
Section 4: Biomed Functions