Section 5: Reference
Pump Alarms
Maximum Infusion
Pressure
..............................
Maximum Time to
Occlusion Alarm
Bolus Volume at
Occlusion Alarm
Pressure
..............................
Power Sources
System Operating
Temperature
58
Low battery power; depleted battery
.....................
power; battery dislodged; pump stopped;
pump fault; low reservoir volume; high
delivery pressure; air in line; disposable
not attached when run attempted; motor
locked; upstream occlusion; reservoir
volume empty; program incomplete; key
stuck; disposable detached.
40.0 psi [2.76 bar]
CADD
...............
siphon valve: 2.0 hours
0.050 ml resolution CADD
tion Sets/Medication Cassette
voirs: <0.15 ml
Two AA alkaline batteries; AC Adapter.
...................
The expected life of two AA batteries is
15 hours at 125 ml/hour, or approxi-
mately 14 days at 10 ml/day (nominal).
This estimate is based on laboratory tests
conducted at room temperature using
two new batteries. Actual battery life will
vary depending on the brand of battery,
battery shelf life, temperature conditions,
and delivery rate. It is recommended that
two new AA batteries be kept available
for replacement if necessary.
An internal battery powers the clock.
When it is depleted, it cannot reliably
maintain the clock time. This battery
must be replaced by the manufacturer.
The internal battery has an expected life
of 5 years.
+2°C to 40°C (35°F to 104°F)
......................
®
Administration Set with anti-
®
Administra-
™
Reser-